Separation of helium and deuterium peaks with a quadrupole mass spectrometer by using the second stability zone in the Mathieu diagram

Separation of helium and deuterium peaks with a quadrupole mass spectrometer by using the second stability zone in the Mathieu diagram
复制标题

使用马蒂厄图中的第二稳定区,使用四极杆质谱仪分离氦峰和氘峰

DOI:
10.1063/1.1143286
复制
发表时间:
1992
影响因子:
1.6
通讯作者:
Y. Murakami
Y. Murakami
中科院分区:
工程技术4区
文献类型:
--
作者:
S. Hiroki;T. Abe;Y. Murakami

文献摘要

被引文献

相似文献

利用Mathieu图中的第二稳定区,用传统的四极杆质谱仪分离了he +和D2+。分离4He+和D2+所需的最小分辨能力为161。在离子能量为14.6 eV的情况下,实验获得了320的分辨率,即半最大全宽度。
4He+ and D2+ have been separated with a conventional quadrupole mass spectrometer by using the second stability zone in the Mathieu diagram. The minimum resolving power needed for the separation of 4He+ and D2+ is 161. A resolution of 320 defined by the full width at half maximum was attained at an ion energy of 14.6 eV in the experiment.