FULLY AUTOMATED MUSCLE ULTRASOUND ANALYSIS (MUSA): ROBUST AND ACCURATE MUSCLE THICKNESS MEASUREMENT
FULLY AUTOMATED MUSCLE ULTRASOUND ANALYSIS (MUSA): ROBUST AND ACCURATE MUSCLE THICKNESS MEASUREMENT
复制标题
DOI:
10.1016/j.ultrasmedbio.2016.08.032
复制
发表时间:
2017-01-01
影响因子:
2.9
通讯作者:
Minetto, Marco Alessandro
中科院分区:
文献类型:
--
作者:
Caresio, Cristina;Salvi, Massimo;Minetto, Marco Alessandro
Musculoskeletal ultrasound imaging allows non-invasive measurement of skeletal muscle thickness. Current techniques generally suffer from manual operator dependency, while all the computer-aided approaches are limited to be semi-automatic or specifically optimized for a single muscle. The aim of this study was to develop and validate a fully automatic method, named MUSA (Muscle UltraSound Analysis), for measurement of muscle thickness on longitudinal ultrasound images acquired from different skeletal muscles. The MUSA algorithm was tested on a database of 200 B-mode ultrasound images of rectus femoris, vastus lateralis, tibialis anterior and medial gastrocnemius. The automatic muscle thickness measurements were compared to the manual measurements obtained by three operators. The MUSA algorithm achieved a 100% segmentation success rate, with mean differences between the automatic and manual measurements in the range of 0.06-0.45 mm. MUSA performance was statistically equal to the operators and its measurement accuracy was independent of the muscle thickness value. (E-mail: filippo.molinari@polito.it) (C) 2016 World Federation for Ultrasound in Medicine & Biology.