Testing of the UV discharge system for LISA Pathfinder
Testing of the UV discharge system for LISA Pathfinder
复制标题
DOI:
10.1063/1.2405047
复制
发表时间:
2006-12
期刊:
影响因子:
--
通讯作者:
P. Wass;L. Carbone;A. Cavalleri;G. Ciani;R. Dolesi;M. Hueller;G. Rochester;M. Schulte;T. Sumner;D. Tombolato;C. Trenkel;S. Vitale;W. Weber
中科院分区:
文献类型:
--
作者:
P. Wass;L. Carbone;A. Cavalleri;G. Ciani;R. Dolesi;M. Hueller;G. Rochester;M. Schulte;T. Sumner;D. Tombolato;C. Trenkel;S. Vitale;W. Weber
We present the results of tests that demonstrate the functionality and performance of the test‐mass discharge system for LISA and LISA Pathfinder. In space, the LISA and LISA Pathfinder test‐masses will accumulate charge from energetic charged particles. Using the photoelectric effect, a system of ultra‐violet (UV) illumination will discharge the test masses. Testing of this UV discharge system, integrated with the LISA Pathfinder engineering model sensor in the torsion pendulum facility at the University of Trento has shown that: isolated test‐masses can be charged positively or negatively using UV light; the rate of charging can be measured and is proportional to the UV power; the test‐mass charge can be continuously held near zero with a combination of light illuminating the test mass and the surrounding sensor housing, without introducing excess force noise above the pendulum detection limit.