A high precision straw tube chamber with cathode readout
A high precision straw tube chamber with cathode readout
复制标题
带阴极读数的高精度吸管管室
DOI:
10.1016/0168-9002(93)91016-g
复制
发表时间:
1993
影响因子:
1.4
通讯作者:
T. Preda
中科院分区:
文献类型:
--
作者:
V. Bychkov;I. Golutvin;Y. Ershov;E. Zubarev;A. Ivanov;V. Lysiakov;A. V. Makhan’kov;S. Movchan;V. Peshekhonov;T. Preda
The high precision straw chamber with cathode readout was constructed and investigated. The 10 mm diameter straws were made of aluminized Mylar with transparent longitudinal window. TheX-coordinate information has been taken from cathode strips as induced charges and investigated with the centroid method. The spatial resolutionσx= 103μm was obtained at a signal-to-noise ratio of about 70. The possible ways to improve the signal-to-noise ratio are discussed.