A Letter to Reconsider the Conditions for Testing Decontaminated N95 Respirators for Emergency Reuse to Address Shortage
A Letter to Reconsider the Conditions for Testing Decontaminated N95 Respirators for Emergency Reuse to Address Shortage
复制标题
关于重新考虑测试经净化的 N95 呼吸器以供紧急再利用以解决短缺问题的信函
DOI:
--
复制
发表时间:
2020
期刊:
影响因子:
--
通讯作者:
Jeonghyeon Ahn
中科院分区:
文献类型:
--
作者:
Eric P. Vejerano;Jeonghyeon Ahn
ABSTRACT
The battle with COVID-19 pandemic has resulted in the shortage of personal protective equipment, particularly, N95 respirators. Healthcare workers who reused N95 respirators may resort to unproven methods of cleaning/sterilization that can severely compromise the respirators’ filtration efficiency. A recently issued guideline will test decontaminated N95 respirators against particles with a median diameter of 0.075 ± 0.020 µm at a flow rate of 85 L min–1. For emergency reuse, these conditions may be too stringent. N95 respirators tested at this flow rate had predicted efficiencies of 0.5 µm). By reconsidering the test conditions, a substantial fraction of used/decontaminated respirators can be reused.