A Real-Time Algorithm to Improve the Response Time of a Clinical Multigas Analyser
A Real-Time Algorithm to Improve the Response Time of a Clinical Multigas Analyser
复制标题
提高临床多种气体分析仪响应时间的实时算法
DOI:
--
复制
发表时间:
1998
影响因子:
2.2
通讯作者:
C. Hahn
中科院分区:
文献类型:
--
作者:
L. Wong;R. Hamilton;E. Palayiwa;C. Hahn
Objective. An algorithm to improve the response time of a clinical respiratory multigas analyser is presented. Methods. The algorithm involves the application of a second order differential equation to the analyser gas output signals in real-time. The adjusted analyser output signals are compared with those of a quadrupole respiratory mass spectrometer sampling and analysing simultaneously. Results. Our results show a close correlation between the adjusted clinical gas analyser and the mass spectrometer signals. Lung volumes derived from a non-invasive sinusoidal inert gas forcing technique, in a model test lung, using the adjusted clinical gas analyser and the mass spectrometer signals demonstrated comparable results. Conclusions. The algorithm provides an improvement on the relatively slow response times of the clinical gas analyser for breath-by-breath time-dependent applications. The same algorithm can also be applied to other instruments which have slow response times.