A polydopamine-modified optical fiber SPR biosensor using electroless-plated gold films for immunoassays
A polydopamine-modified optical fiber SPR biosensor using electroless-plated gold films for immunoassays
复制标题
聚多巴胺修饰光纤 SPR 生物传感器,使用化学镀金膜进行免疫分析
DOI:
10.1016/j.bios.2015.06.080
复制
发表时间:
2015-12-15
影响因子:
12.6
通讯作者:
He, Zhimin
中科院分区:
文献类型:
--
作者:
Shi, Se;Wang, Libing;He, Zhimin
A sensitive and stable electroless-plated gold film for the preparation of an optical fiber surface plasmon resonance (SPR) sensor is presented in this work, together with a facile antibody immobilization method. Gold nanoparticles were uniformly adsorbed onto the surface of an optical fiber forming a film with a thickness of approximately 563 nm. The sensor had a high sensitivity with 2054 nm/RIU and 3980 nm/RIU in the refractive index ranges of 1.333-1.359 and 1.359-1386, respectively. An SPR biosensor was developed based on polydopamine-modified gold film (PDA-Au), which was fabricated by a simple and quick spontaneous polymerization of dopamine (DA) on the gold film. When goat anti-human IgG antibodies were immobilized, the FDA-Au surface had a larger resonant wavelength shift of 66.21 nm compared with the traditional 11-mercaptoundecanoic acid-modified gold film (MUA-Au) surface. In addition, the PDA-Au surface enabled the sensitive and selective determination of human IgG down to a concentration of 2 mu g mL(-1) with a high sensitivity of 0.41 nm per mu g mL(-1). The FDA-Au surface exhibited an approximately four fold higher sensitivity and an about seven fold lower LOD than the MUA-Au surface to human IgG. (C) 2015 Elsevier B.V. All rights reserved.