Roller screw electric motor ventricular assist device.
Roller screw electric motor ventricular assist device.
复制标题
滚柱丝杠电动马达心室辅助装置。
DOI:
--
复制
发表时间:
1986
期刊:
影响因子:
--
通讯作者:
Pierce Ws
中科院分区:
文献类型:
--
作者:
Wayne E. Richenbacher;Walter E. Pae;James A. Magovern;Gerson Rosenberg;Alan J. Snyder;Pierce Ws
The roller screw electric VAD is easier to manufacture and 25% lighter than the previously described drum cam model. This device requires 12 to 15 W to pump 6 to 8 l/min with minimal hemolysis. The motor drive has functioned for periods up to 93 days in vivo with no measurable wear. The compliance chamber volume varies by 100 cc during VAD function but does so while maintaining pressure variations below 15 mmHg. Compliance chamber volume loss of 2 to 5 cc/day is explained by gas transport through SPU. The subcutaneous sampling port provides ready access to the sealed system.