Mechanical assessment of proprietary and improvised pelvic binders for use in the prehospital environment.

Mechanical assessment of proprietary and improvised pelvic binders for use in the prehospital environment.
复制标题

对院前环境中使用的专有和简易骨盆固定器进行机械评估。

DOI:
10.1136/military-2023-002398
复制
发表时间:
2023
影响因子:
1.5
通讯作者:
Howe TJ
Howe TJ
中科院分区:
医学4区
文献类型:
--
作者:
Howe TJ

文献摘要

相似文献

骨盆骨折常由高能量创伤引起,死亡率高达10%,并发症严重。骨盆固定器应用于疑似骨盆损伤,以稳定骨折,减少出血和加强填塞。捆绑器必须以足够的力固定骨盆,以达到这种效果。本研究的目的是量化专有和即兴骨盆绑定器的能力,以保持一个目标的张力随着时间的推移。MethodsThe三个专有和三个即兴绑定器的能力,以保持约束力为2小时进行了测试。使用单轴材料试验机将每种粘合剂拉伸至150 N,然后保持位移2小时;记录每种粘合剂随时间的张力下降。保持张力高于130 N的能力,2小时后被设置为度量的粘合剂performance.ResultsThe中位数张力在2小时以上的SAM骨盆吊带II和T-POD骨盆稳定装置,并低于130 N的普罗米修斯骨盆夹板,现场权宜骨盆夹板(FES)和个人服装系统,多地形模式战斗裤粘合剂。2小时后,简易FES中的张力约为目标130 N;然而,在40%的测试中,它保持在130 N以上。结论在2小时的测试期间,固定器维持足够张力以治疗骨盆骨折的能力各不相同。FES在我们的测试制度下表现良好;由于成本和重量相对较低,它是严峻环境下专有粘合剂的良好替代品。
IntroductionPelvic fractures often result from high-energy trauma and are associated with a 10% mortality rate and significant morbidity. Pelvic binders are applied in suspected pelvic injury to stabilise fractured bone, decrease bleeding and potentiate tamponade. A binder must hold the pelvis with sufficient force for this effect to be achieved. This study aims to quantify the ability of proprietary and improvised pelvic binders to hold a target tensile force over time.MethodsThe ability of three proprietary and three improvised binders to hold a binding force for 2 hours was tested. A uniaxial materials testing machine was used to tension each binder to 150 N and then hold the displacement for 2 hours; the drop in tension over time was recorded for each binder. The ability to hold tension above 130 N after 2 hours was set as the metric of binder performance.ResultsThe median tension at 2 hours was above 130 N for the SAM Pelvic Sling II and T-POD Pelvic Stabilisation Device and was below 130 N for the Prometheus Pelvic Splint, field-expedient pelvic splint (FES) and the Personal Clothing System-Multi-Terrain Pattern Combat Trousers binders. The tension in the improvised FES after 2 hours was approximately at the target 130 N; however, in 40% of the tests, it held above 130 N.ConclusionsBinders varied in their ability to maintain sufficient tension to treat a pelvic fracture over the 2-hour testing period. The FES performed well under our testing regime; with relatively low cost and weight, it represents a good alternative to proprietary binders for the austere environment.