A magnification mammography technique using storage phosphor plates: Experimental investigations concerning spatial resolution and detection of microcalcifications.

A magnification mammography technique using storage phosphor plates: Experimental investigations concerning spatial resolution and detection of microcalcifications.
复制标题

DOI:
10.1055/s-2007-1015512
复制
发表时间:
1997-08-01
影响因子:
1.8
通讯作者:
Grabbe, E
Grabbe, E
中科院分区:
医学4区
文献类型:
--
作者:
Funke, M;Hermann, KP;Grabbe, E

文献摘要

被引文献

相似文献

目的:由于常规使用的存储荧光系统的乳房X线摄影已被其不足的空间分辨率为5线对/mm的限制,一个组合的放大乳房X线摄影技术与存储荧光板进行了研究,以检测微钙化。材料和方法:一种新的乳腺X线摄影系统,带有一个使用0.05-0.12 mm阳极的微聚焦管,允许以1.7倍放大率(m)获得乳房的调查视图,并以4倍放大率获得点视图。数字图像接收器包括高分辨率存储荧光板。为了确定空间分辨率,获得对比度传递曲线,并通过ROC(受试者工作特征)分析研究微钙化的检测。结果如下:空间分辨率为数字调查的意见(m=1.7)是8线对/毫米和点视图(m=4)是18线对/毫米。ROC分析表明,数字放大技术的性能显着高于传统的屏幕电影乳腺摄影技术,结论:数字乳腺摄影的局限性,相对于空间分辨率可以克服使用高放大率技术。
Purpose: Since the routine use of storage phosphor systems for mammography has been limited by its inadequate spatial resolution of 5 linepairs/mm, a combination of a magnification mammography technique with storage phosphor plates was investigated to detect microcalcifications. Material and Methods: A new mammography system with a microfocus tube using an anode of 0.05-0.12 mm allowed to obtain survey views of the breast with 1.7x magnification (m), and spot views with 4x magnification. The digital image receptor comprised a high resolution storage phosphor plate. To determine spatial resolution, contrast transfer curves were obtained, and the detection of microcalcifications was investigated by ROC (receiver operating characteristic) analysis. Results: Spatial resolution for digital survey views (m=1.7) was 8 linepairs/mm and for spot views (m=4) was 18 linepairs/mm. ROC analysis demonstrated a significantly higher performance of the digital magnification technique compared to the conventional screen-film mammography technique, Conclusions: The limitations of digital mammography with respect to spatial resolution can be overcome by using a high magnification technique.