Validation of the AM5600 ambulatory blood pressure monitor in children and adolescents.

Validation of the AM5600 ambulatory blood pressure monitor in children and adolescents.
复制标题

DOI:
10.1097/mbp.0b013e3283102cfe
复制
发表时间:
2008-12
影响因子:
1.3
通讯作者:
Alpert BS
Alpert BS
中科院分区:
医学4区
文献类型:
--
作者:
Jones DP;Richey PA;Alpert BS

文献摘要

相似文献

我们使用AM5600测量儿童和青少年的动态血压,以评估血压与心血管疾病危险因素的关系。虽然这种监护仪的使用在成人中已有报道,但尚未在儿科受试者中得到验证。在这项研究中,我们评估了在7-18岁的儿童中,与水银血压计相比,监护仪的准确性。我们发现监测器与汞控装置的收缩压和舒张压差均值分别为0.29±3.5和0.045±3.7 mmHg,符合AAMI对设备使用的标准。根据英国高血压协会(British Hypertension Society)的数据,两种设备之间的累积读数百分比不同,给该设备打了a级。
We measured ambulatory blood pressure using the AM5600 in children and adolescents participating in a research study to assess the relationship of BP to risk factors for cardiovascular disease. Although this use of this monitor has been previously reported in adults, it has not been validated in pediatric subjects. In this study, we assess the accuracy of the monitor as compared to the mercury sphygmomanometer in children ages 7-18 years of age. We found that the mean of the difference between the monitor and the mercury device was 0.29 ± 3.5 and 0.045 ± 3.7 mmHg for systolic and diastolic BP respectively, which fulfills the AAMI standard for use of a device. The cumulative percent of readings between the two devices which differed assigned the device a grade of A according to the British Hypertension Society..