Accuracy and reliability of a subcutaneous continuous glucose monitoring device in critically ill patients
Accuracy and reliability of a subcutaneous continuous glucose monitoring device in critically ill patients
复制标题
DOI:
10.1007/s10877-017-0086-z
复制
发表时间:
2018-10-01
影响因子:
2.2
通讯作者:
van der Voort, P. H. J.
中科院分区:
文献类型:
--
作者:
Rijkenberg, S.;van Steen, S. C.;van der Voort, P. H. J.
Subcutaneous continuous glucose monitoring (CGM) may have benefits in achieving glycemic control in critically ill patients. The aim of this study was to assess the accuracy and reliability of the FreeStyle Navigator I in critically ill patients and to assess patient related factors influencing the accuracy and reliability. This study is a retrospective analysis of data from a randomized controlled trial conducted in a 20-bed mixed intensive care unit. Analytical accuracy, clinical accuracy and reliability were assessed against arterial blood glucose samples as reference. Assessment was according to recent consensus recommendations with median absolute relative difference (median ARD), Bland-Altman plots, the ISO system accuracy standards (ISO 15197:2013) and Clarke error grid analysis (CEG). We analyzed 2840 paired measurements from 155 critically ill patients. The median ARD of all paired values was 13.3 [6.9-22.1]%. The median ARD was significantly higher in both the hypoglycemic and the hyperglycemic range (32.4 [12.1-53.4] % and 18.7 [10.7-28.3] % respectively, p