Spiral MR fingerprinting at 7 T with simultaneous B1 estimation
Spiral MR fingerprinting at 7 T with simultaneous B1 estimation
复制标题
DOI:
10.1016/j.mri.2017.04.003
复制
发表时间:
2017-09-01
影响因子:
2.5
通讯作者:
Tosetti, Michela
中科院分区:
文献类型:
--
作者:
Buonincontri, Guido;Schulte, Rolf F.;Tosetti, Michela
Magnetic resonance fingerprinting is an efficient, new approach for quantitative imaging with MR. We aimed to extend this technique to cases with B1+ inhomogeneities within the imaging volume.Previous approaches have used abrupt changes in flip angles to estimate the B1+ field simultaneously with T1 and T2, using a Cartesian approach in a small-animal scanner at 4.7T. Here, we evaluated whether a similar approach would be suitable for imaging human brains using spiral readouts with a 7T scanner.We found that our modified scheme could significantly reduce the adverse effects of B1+ inhomogeneities even in extreme cases, reducing both the bias and the variance in T2 estimations by an order of magnitude when compared to literature methods. Acquisitions used less than 1.5W/kg SAR and could be performed in 12s per slice. In conclusion, our approach can be used to perform quantitative imaging of the brain at 7T in a short time, simultaneously estimating the B1+ profile. (C) 2017 Elsevier Inc. All rights reserved.