Influences of tissue absorption and scattering on diffuse correlation spectroscopy blood flow measurements.
Influences of tissue absorption and scattering on diffuse correlation spectroscopy blood flow measurements.
复制标题
DOI:
10.1364/boe.2.001969
复制
发表时间:
2011-07-01
影响因子:
3.4
通讯作者:
Yu G
中科院分区:
文献类型:
--
作者:
Irwin D;Dong L;Shang Y;Cheng R;Kudrimoti M;Stevens SD;Yu G
In this study we evaluate the influences of optical property assumptions on near-infrared diffuse correlation spectroscopy (DCS) flow index measurements. The optical properties, absorption coefficient (µa) and reduced scattering coefficient (µs′), are independently varied using liquid phantoms and measured concurrently with the flow index using a hybrid optical system combining a dual-wavelength DCS flow device with a commercial frequency-domain tissue-oximeter. DCS flow indices are calculated at two wavelengths (785 and 830 nm) using measured µa and µs′ or assumed constant µa and µs′. Inaccurate µs′ assumptions resulted in much greater flow index errors than inaccurate µa. Underestimated/overestimated µs′ from −35%/+175% lead to flow index errors of +110%/−80%, whereas underestimated/overestimated µa from −40%/+150% lead to −20%/+40%, regardless of the wavelengths used. Examination of a clinical study involving human head and neck tumors indicates up to +280% flow index errors resulted from inter-patient optical property variations. These findings suggest that studies involving significant µa and µs′ changes should concurrently measure flow index and optical properties for accurate extraction of blood flow information.