Sagittal laser optical tomography for imaging of rheumatoid finger joints
Sagittal laser optical tomography for imaging of rheumatoid finger joints
复制标题
DOI:
10.1088/0031-9155/49/7/005
复制
发表时间:
2004-04-07
影响因子:
3.5
通讯作者:
Beuthan, J
中科院分区:
文献类型:
--
作者:
Hielscher, AH;Klose, AD;Beuthan, J
We present a novel optical tomographic imaging system that was designed to determine two-dimensional spatial distribution of optical properties in a sagittal plane through finger joints. The system incorporates a single laser diode and a single silicon photodetector into a scanning device that records spatially resolved light intensities as they are transmitted through a finger. These data are input to a model-based iterative image reconstruction (MOBIIR) scheme, which uses the equation of radiative transfer (ERT) as a forward model for light propagation through tissue. We have used this system to obtain tomographic images of six proximal interphalangeal finger joints from two patients with rheumatoid arthritis. The optical images were compared to clinical symptoms and ultrasound images.