Pre-operative transthoracic real-time three-dimensional echocardiography in patients undergoing mitral valve repair: accuracy in cases with simple vs. complex prolapse lesions
Pre-operative transthoracic real-time three-dimensional echocardiography in patients undergoing mitral valve repair: accuracy in cases with simple vs. complex prolapse lesions
复制标题
DOI:
10.1093/ejechocard/jeq066
复制
发表时间:
2010-10-01
影响因子:
--
通讯作者:
Pepi, Mauro
中科院分区:
文献类型:
--
作者:
Tamborini, Gloria;Muratori, Manuela;Pepi, Mauro
The aim of this study, undertaken in patients who underwent mitral valve (MV) repair surgery, was to evaluate the accuracy of pre-operative three-dimensional (3D) transthoracic echocardiography (TTE) in the evaluation of MV pathology in cases with simple or complex lesions.Two hundred consecutive patients with severe mitral regurgitation due to degenerative MV prolapse underwent a complete 3DTTE the day before surgery. Three-dimensional TTE data were compared with MV surgical inspection. Three-dimensional echocardiography was feasible in a relatively short time (5 +/- 3 min) with good (67%) and optimal (21%) imaging quality in the majority of cases. Three-dimensional TTE allowed an accurate identification (95% accuracy) of all MV lesions. Seventy-three (36.5%) patients had simple lesions at 3DTTE and 71 of them (97.2%) underwent a simple surgical procedure; 127 (63.5%) had complex lesions at 3DTTE and, in these cases, surgeons performed either simple procedures (48%) or complex procedures (47.2%) or valve replacement in 4.7% (after a first attempt for repair).Three-dimensional TTE is feasible, not time-consuming, and accurate in identifying cases with simple vs. complex MV lesions.