Effects of Ti/TiN multilayer on corrosion resistance of nickel–titanium orthodontic brackets in artificial saliva
Effects of Ti/TiN multilayer on corrosion resistance of nickel–titanium orthodontic brackets in artificial saliva
复制标题
DOI:
10.1016/j.corsci.2007.03.041
复制
发表时间:
2007-10
影响因子:
8.3
通讯作者:
Chenglong Liu;P. Chu;Guoqiang Lin;Dazhi Yang
中科院分区:
文献类型:
--
作者:
Chenglong Liu;P. Chu;Guoqiang Lin;Dazhi Yang
The effects of multilayered Ti/TiN or single-layered TiN films deposited by pulse-biased arc ion plating (PBAIP) on the corrosion behavior of NiTi orthodontic brackets in artificial saliva are investigated. The multilayered Ti/TiN coating is found to exhibit a greater free corrosion potential, much lower passive current density, and no breakdown up to 1.5V. Moreover, electrochemical impedance spectroscopy (EIS) results indicate that the multilayered Ti/TiN coating has a larger impedance and lower porosity which is believed to be responsible for the exceedingly low metal ion release rate during 720h exposure in the test solution. Visual inspection of the surfaces reveals different corrosion processes for the TiN and multilayered Ti/TiN coatings.