A power recovery strategy dedicated to implantable applications
A power recovery strategy dedicated to implantable applications
复制标题
专用于植入式应用的功率恢复策略
DOI:
10.1109/icecs.2003.1301731
复制
发表时间:
2003
期刊:
影响因子:
--
通讯作者:
M. El
中科院分区:
文献类型:
--
作者:
Yamu Hu;M. Sawan;M. El
We present in this paper a power recovery strategy dedicated to the biotelemetry application of cortical/nerve stimulation. The proposed strategy allows providing dual regulated output voltages 3.3V/1.8V to the electrodes driver and other implantable circuitry, respectively. Its low dropout voltage makes a high power efficiency attainable. One fabricated integrated prototype in a CMOS 0.18 /spl mu/m technology has demonstrated its feasibility, providing a load current driving ability of 5 mA for each one of the two supply voltages.