Using cerebral oximetry to prevent adverse outcomes during cardiac surgery

Using cerebral oximetry to prevent adverse outcomes during cardiac surgery
复制标题

DOI:
10.1177/0267659110393298
复制
发表时间:
2011-03-01
期刊:
影响因子:
1.2
通讯作者:
Tang, A.
Tang, A.
中科院分区:
医学4区
文献类型:
--
作者:
Faulkner, J. T.;Hartley, M.;Tang, A.

文献摘要

被引文献

相似文献

我们报告了一例脑氧合减少的病例,如果常规使用脑血氧测定法未检测到,则可能导致不良结局。该病例研究表明,上级腔静脉内单级静脉插管的意外重新调整仅导致脑血氧测定装置警告潜在问题。所有其他监测装置都保持在标准操作参数范围内,在整个事件期间没有出现偏差。
We report a case of reduced cerebral oxygenation which had the possibility of leading to an adverse outcome if it had not been detected by the routine use of cerebral oximetry. This case study illustrates that an inadvertent re-adjustment of a single-stage venous cannula within the superior vena cava resulted only in the cerebral oximetry device alerting to a potential problem. All other monitoring devices remained within standard operating parameters, with no deviation throughout the duration of the incident.