Human exposure to Bisphenol A and liver health status: Quantification of urinary and circulating levels by LC-MS/MS
Human exposure to Bisphenol A and liver health status: Quantification of urinary and circulating levels by LC-MS/MS
复制标题
DOI:
10.1016/j.jpba.2017.02.058
复制
发表时间:
2017-06-05
影响因子:
3.4
通讯作者:
Diano, Nadia
中科院分区:
文献类型:
--
作者:
Nicolucci, Carla;Erric, Sonia;Diano, Nadia
A selective and highly sensitive analytical methodology for determination of Bisphenol A in human plasma was developed and validated. The method was based on selective liquid/solid extraction, combined with liquid chromatography-electrospray ionization tandem mass spectrometry in the multiple reaction monitoring mode and negative ionization. The linearity of the detector response was verified in human plasma over the concentration range 0.100-200 ng mL(-1). The detection limit was 0.03 ng mL(-1) and the quantification limit was 0.100 ng mL(-1). The analytical features of the proposed in-house validated method were satisfactory: precision was