Validation of a three-dimensional facial scanning system based on structured light techniques

Validation of a three-dimensional facial scanning system based on structured light techniques
复制标题

基于结构光技术的三维面部扫描系统的验证

DOI:
10.1016/j.cmpb.2009.01.010
复制
发表时间:
2009-06-01
影响因子:
6.1
通讯作者:
Lin, Jiuxiang
Lin, Jiuxiang
中科院分区:
工程技术2区
文献类型:
--
作者:
Ma, Lili;Xu, Tianmin;Lin, Jiuxiang

文献摘要

被引文献

相似文献

The aim of this study was to validate a newly developed three-dimensional (3D) structured light scanning system in recording the facial morphology. The validation was performed in three aspects including accuracy, precision and reliability. The accuracy and precision were investigated using a plaster model with 19 marked landmarks. The accuracy was determined by comparing the coordinates from the 3D images and from the coordinates measure machine (CMM). The precision was quantified through the repeated landmarks location on 3D images. The reliability was investigated in 10 adult volunteers. Each was scanned five times in 3 weeks. The 3D images acquired at different times were compared with each other to measure the reliability. We found that the accuracy was 0.93 mm, the precision was 0.79 mm, the reliability was 0.2 mm. These findings suggested that the structured light scanning system was accurate, precise and reliable to record the facial morphology for both clinic and research purposes. (C) 2009 Elsevier Ireland Ltd. All rights reserved.