Proton beam therapy interference with implanted cardiac pacemakers
Proton beam therapy interference with implanted cardiac pacemakers
复制标题
DOI:
10.1016/j.ijrobp.2008.01.062
复制
发表时间:
2008-11-01
影响因子:
7
通讯作者:
Tokuuye, Koichi
中科院分区:
文献类型:
--
作者:
Oshiro, Yoshiko;Sugahara, Shinji;Tokuuye, Koichi
Purpose: To investigate the effect of proton beam therapy (PBT) on implanted cardiac pacemaker function.Methods and Materials: After a phantom study confirmed the safety of PBT in patients with cardiac pacemakers, we treated 8 patients with implanted pacemakers using PBT to a total tumor dose of 33-77 gray equivalents (GyE) in dose fractions of 2.2-6.6 GyE. The combined total number of PBT sessions was 127. Although all pulse generators remained outside the treatment field, 4 patients had pacing leads in the radiation held. All patients were monitored by means of electrocardiogram during treatment, and pacemakers were routinely examined before and after PBT.Results: The phantom study showed no effect of neutron scatter oil pacemaker generators. In the study changes in heart rate occurred three times (2.4%) in 2 patients. However, these patients remained completely asymptomatic throughout the PBT course.Conclusions: PBT can result in pacemaker malfunctions that manifest as changes in pulse rate and pulse patterns. Therefore. patients with cardiac pacemakers should be monitored by means of electrocardiogram during PBT. (C) 2008 Elsevier Inc.