Characteristics of a new fully programmable blood sampling device for monitoring blood radioactivity during PET
Characteristics of a new fully programmable blood sampling device for monitoring blood radioactivity during PET
复制标题
DOI:
10.1007/s002590000405
复制
发表时间:
2001-01-01
期刊:
影响因子:
--
通讯作者:
Lammertsma, AA
中科院分区:
文献类型:
--
作者:
Boellaard, R;van Lingen, A;Lammertsma, AA
The first performance tests of a new fully programmable blood sampling device for monitoring blood radioactivity during positron emission tomography (PET) are described. Blood is withdrawn through 1-mm internal diameter tubing using an infusion pump which can be operated at rates varying from 0 to 600 ml/h. Activity in blood is measured by a 6-cm-thick bismuth germanate crystal connected to a photomultiplier tube and multichannel analyser (MCA) which are positioned within 6 cm lead shielding. Positioning of the tubing is an exact and simple procedure. The minimal readout time of the MCA is 1 s. Two independent energy windows can be set. Operation of the pump and MCA is fully controlled by a PC, i.e. sampling time, interval time and pump rate can be varied at any time during the PET scan by user-defined scripts. A number of characteristics of the new system were studied, such as sensitivity, dead time, linearity, effect of background radiation and pump rate as a function of input pressure. In addition, dispersion was measured as a function of pump rate. Finally, first clinical results were compared with manual samples. The sensitivity equalled 0.7 and 0.2 cps/Bq for 511- and 1022-keV 30% energy windows, respectively, and the system dead time was 500 ns. The system remained linear within 2% with activity concentrations up to 2.5 MBq/cc. Short-term reproducibility was better than 3% for a 1-h period. Long-term reproducibility was about 5% (1SD), which was mainly caused by variation in the diameter of the tubing. If the device was positioned in such a way that maximum shielding was directed towards the patient, the effects of background radiation from the patient on the measured activity concentration for clinically relevant conditions was minimal (