Filling the dead-time gap in zero echo time MRI: Principles compared
Filling the dead-time gap in zero echo time MRI: Principles compared
复制标题
DOI:
10.1002/mrm.26875
复制
发表时间:
2018-04-01
影响因子:
3.3
通讯作者:
Pruessmann, Klaas P.
中科院分区:
文献类型:
--
作者:
Froidevaux, Romain;Weiger, Markus;Pruessmann, Klaas P.
PurposeMRI of tissues with short coherence lifetimes T-2 or T2* can be performed efficiently using zero echo time (ZTE) techniques such as algebraic ZTE, pointwise encoding time reduction with radial acquisition (PETRA), and water- and fat-suppressed proton projection MRI (WASPI). They share the principal challenge of recovering data in central k-space missed due to an initial radiofrequency dead time. The purpose of this study was to compare the three techniques directly, with a particular focus on their behavior in the presence of ultra-short-lived spins.MethodsThe most direct comparison was enabled by aligning acquisition and reconstruction strategies of the three techniques. Image quality and short-T2* performance were investigated using point spread functions, 3D simulations, and imaging of phantom and bone samples with short (