Retinal Blood Flow Response to Hyperoxia Measured With En Face Doppler Optical Coherence Tomography.

Retinal Blood Flow Response to Hyperoxia Measured With En Face Doppler Optical Coherence Tomography.
复制标题

DOI:
10.1167/iovs.15-18917
复制
发表时间:
2016-07-01
影响因子:
4.4
通讯作者:
Huang D
Huang D
中科院分区:
医学2区
文献类型:
--
作者:
Pechauer AD;Tan O;Liu L;Jia Y;Hou V;Hills W;Huang D

文献摘要

被引文献

相似文献

To use multiplane en face Doppler optical coherence tomography (OCT) to measure the change in total retinal blood flow (TRBF) in response to hyperoxia. One eye of each healthy human participant (n = 8) was scanned with a commercial high-speed (70-kHz) spectral OCT system. Three repeated scans were captured at baseline and after 10 minutes of oxygen (hyperoxia) by open nasal mask. The procedure was performed twice on day 1 and once more on day 2. Blood flow of each vein was estimated using Doppler OCT at an optimized en face plane. The TRBF was summed from all veins at the optic disc. The TRBF hyperoxic response was calculated as the TRBF percent change from baseline. Participants experienced a 23.6% ± 10.7% (mean ± standard deviation [SD]) decrease (P < 0.001, paired t-test) in TRBF during hyperoxia. The within-day repeatability of baseline TRBF was 4.1% and the between-day reproducibility was 10.9% coefficient of variation (CV). Between-grader reproducibility was 3.9% CV. The repeatability and reproducibility (pooled SD) of hyperoxic response were 6.1% and 6.4%, respectively. The multiplane en face Doppler OCT algorithm was able to detect, in all participants, a decreased TRBF in response to hyperoxia. The response magnitude for each participant varied among repeated trials, and the averaging of multiple trials was helpful in establishing the individual response. This technique shows good potential for the clinical investigation of vascular autoregulation.