Impact of scale standardization on images of digital radiography systems.
Impact of scale standardization on images of digital radiography systems.
复制标题
尺度标准化对数字放射线照相系统图像的影响。
DOI:
--
复制
发表时间:
1997
期刊:
影响因子:
--
通讯作者:
P. van der Stelt
中科院分区:
文献类型:
--
作者:
C. Versteeg;G. Sanderink;W. Geraets;P. van der Stelt
OBJECTIVES
To test the effect of altering image size on diagnostic quality.
METHODS
Endodontic files, size 10 and 15 were prepared to full root length and 1.5 mm short, in upper and lower molars and premolars. Digital images of the Sidexis (S) and Visualix/Vixa (V) system were reduced in size to Digora (D) scale, and the D images were enlarged to S scale. ROC analysis was performed and data analysed with MANOVA statistics.
RESULTS
Assessment of root canal length with file size 10 was less accurate for the S images reduced in size, compared with the original S images (P < 0.011). All other differences were not significant (P > 0.081).
CONCLUSIONS
Relevant diagnostic information may be lost when images are reduced in size. Therefore, for optimal presentation, the smaller images should be enlarged rather than the larger ones reduced.
影响因子:
1.3
作者:
Schreiner,S;Paschal,CB;Galloway,RL
通讯作者:
Galloway,RL