Imaging performance of a prototype scanned-slot digital mammography system
Imaging performance of a prototype scanned-slot digital mammography system
复制标题
原型扫描槽数字乳腺X线摄影系统的成像性能
DOI:
10.1117/12.154579
复制
发表时间:
1993
期刊:
影响因子:
--
通讯作者:
B. Starkoski
中科院分区:
文献类型:
--
作者:
Andrew D. A. Maidment;M. Yaffe;D. Plewes;G. Mawdsley;I. Soutar;B. Starkoski
A prototype of a clinical scanned-slot digital mammography imaging system has been developed, which demonstrates better contrast sensitivity and latitude than current state-of-the- art film-screen mammography systems. The detector consists of a Gd2O2S:Tb phosphor screen coupled via a 2-to-1 demagnifying fiber-optic taper to two time-delay integration (TDI) charge-coupled device (CCD) image arrays. Images are obtained by scanning the 4.0 mm wide by 21 cm long detector across the image field. An 18 cm by 21 cm image contains 2900 by 4032 pixels, of dimension 62 micrometers X 52 micrometers at the detector. Currently, images are produced in 7.8 seconds using a 40 kV tungsten-target spectrum with a total heat load of 50 kJ, giving a mean glandular dose of 0.85 mGy (85 mrad) to a 5 cm thick 50% glandular, 50% adipose breast. The detector has a limiting resolution of 9.5 lp/mm. A clinical version of this prototype, which incorporates several improvements, is being constructed.