Two-dimensional position resolution and correction of position dependence of pulse heights of superconducting series-junction detectors
Two-dimensional position resolution and correction of position dependence of pulse heights of superconducting series-junction detectors
复制标题
超导串联结探测器脉冲高度的二维位置分辨率和位置依赖性校正
DOI:
10.1016/s0168-9002(03)01370-6
复制
发表时间:
2003
影响因子:
1.4
通讯作者:
K. Taniguchi
中科院分区:
文献类型:
--
作者:
M. Kurakado;S. Kamihirata;A. Kagamihata;K. Hirota;H. Hashimoto;Hiromi Sato;H. Hotchi;H. Shimizu;K. Taniguchi
One of the drawbacks of the series-junction detector is that signal heights depend on position of incident photon. A series-junction detector with two-dimensional position resolution has been developed. The effective area is 1.5×1.5mm2. The detector was irradiated by α particles and a position resolution of about 50μm was obtained. Making use of this position resolution, the position dependence of pulse heights can be corrected, such that the energy resolution was improved from 11% to 0.79%.