In Vitro Tests for Aerosol Deposition II: IVIVCs for Different Dry Powder Inhalers in Normal Adults
In Vitro Tests for Aerosol Deposition II: IVIVCs for Different Dry Powder Inhalers in Normal Adults
复制标题
DOI:
10.1089/jamp.2012.0975
复制
发表时间:
2013-06-01
影响因子:
3.4
通讯作者:
Byron, Peter R.
中科院分区:
文献类型:
--
作者:
Delvadia, Renishkumar;Hindle, Michael;Byron, Peter R.
Background: A new in vitro test method for dry powder inhalers (DPIs) was recently found to be predictive of the published in vivo results for Budelin Novolizer. The present study was intended to assess the method's robustness by evaluating correlations between average drug deposition in vitro and in vivo from five different DPIs.Methods: In vitro drug deposition from five marketed DPIs was assessed in a realistic physical airway model of a "medium" sized adult in an experimental setup that allowed deposition to be characterized regionally for carefully selected simulated air flow rate versus time profiles. The DPIs studied were Spiriva (R) HandiHaler (R), Relenza (R) Diskhaler (R), Salbutamol Easyhaler (R), Pulmicort (R) Turbuhaler (R), and Foradil (R) Aerolizer (R). In vitro regional deposition results were compared with those reported in the literature in order to create in vitro-in vivo correlations (IVIVCs) for each inhaler.Results: Mean percent total lung deposition (TLD-SD) in vitro for Spiriva HandiHaler, Relenza Diskhaler, Salbutamol Easyhaler, Pulmicort Turbuhaler, and Foradil Aerolizer were 17.3 +/- 1.2, 22.6 +/- 1.1, 29.0 +/- 1.1, 28.0 +/- 3.0, and 21.7 +/- 1.2, respectively. These results showed excellent agreement with reported in vivo values, with absolute prediction errors in TLD of