System Modeling and Evaluation of a Prototype Inverted-Compound Eye Gamma Camera for the Second Generation MR Compatible SPECT.

System Modeling and Evaluation of a Prototype Inverted-Compound Eye Gamma Camera for the Second Generation MR Compatible SPECT.
复制标题

用于第二代 MR 兼容 SPECT 的原型倒置复眼伽玛相机的系统建模和评估。

DOI:
10.1016/j.nima.2019.04.001
复制
发表时间:
2020
期刊:
Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment
影响因子:
--
通讯作者:
Meng,Ling-Jian
Meng,Ling-Jian
中科院分区:
--
文献类型:
--
作者:
Lai,Xiaochun;Zannoni,ElenaM;George,Jonathan;Meng,Ling-Jian

文献摘要

相似文献

We have reported the design of the MRC-SPECT-II system based on the inverted-compound-eye (ICE) gamma camera to offer a> 1% detection efficiency while maintaining a sub-500 μ m imaging resolution (Lai and Meng, 2018). One of the key challenges of using the ICE camera for SPECT imaging is whether one could develop an accurate point response function (PRF), given its complex aperture design and low fractionation accuracy of 3D printing. In this work, we will discuss (I) a combined experimental and analytical approach for deriving the precise PRF, and (II) an experimental imaging study to demonstrate the feasibility of using the ICE-camera for acquiring high-quality SPECT images with a sub-500 μ m resolution. These studies would help to overcome one of the major hurdles for implement ICE-cameras for practical SPECT imaging.