Comparison of IgE values as determined by different solid phase radioimmunoassay methods
Comparison of IgE values as determined by different solid phase radioimmunoassay methods
复制标题
不同固相放射免疫分析方法测定IgE值的比较
DOI:
--
复制
发表时间:
1976
期刊:
影响因子:
--
通讯作者:
N. Kjellman
中科院分区:
文献类型:
--
作者:
S. Johansson;Asta Berglund;N. Kjellman
Accurate measurement of low IgE concentrations is technically difficult. In this paper results obtained by a direct sandwich and three inhibition methods of radioimmunoassays are compared. For values above 50 U/ml good correlation was obtained with all methods. Below 50 U/ml, however, the inhibition methods tended to yield falsely high values. For very low concentrations, 1–10 U/ml the best correlation was obtained between the direct sandwich test (PRIST*) and the inhibition test using a correction factor to allow for the non‐specific effect of serum.