A new compact aerosol concentrator for use in conjunction with low flow-rate continuous aerosol instrumentation
A new compact aerosol concentrator for use in conjunction with low flow-rate continuous aerosol instrumentation
复制标题
DOI:
10.1016/j.jaerosci.2004.11.015
复制
发表时间:
2005-08-01
影响因子:
4.5
通讯作者:
Sioutas, C
中科院分区:
文献类型:
--
作者:
Geller, MD;Biswas, S;Sioutas, C
Currently available versatile aerosol concentration enrichment systems (VACES) have proven useful for providing elevated levels of ambient particulate matter to human and animal exposures, as well as for the collection of particles in aqueous solutions for in vitro studies. Previous studies have demonstrated that such systems do not significantly alter the physical or chemical properties of the particles. The current VACES configuration consumes significant electrical power for pumping and cooling, and requires attended operation by expert operators. A recent application of the VACES has been to provide a concentrated aerosol stream to continuous particle mass spectrometers in order to increase the spectrometer's hit rate or sensitivity. These instruments usually require low intake flow rates (