Measurement error in frequency measured using wavelength meter due to residual moisture in interferometer and a simple method to avoid it
Measurement error in frequency measured using wavelength meter due to residual moisture in interferometer and a simple method to avoid it
复制标题
由于干涉仪中残留水分而导致使用波长计测量频率的测量误差以及避免该误差的简单方法
DOI:
10.1088/0957-0233/27/11/115004
复制
发表时间:
2016
影响因子:
2.4
通讯作者:
H. Abe
中科院分区:
文献类型:
--
作者:
K. Hashiguchi;H. Abe
We have experimentally evaluated the accuracy of the frequency measured using a commonly used wavelength meter in the near-infrared region, which was calibrated in the visible region. An error of approximately 50 MHz was observed in the frequency measurement using the wavelength meter in the near-infrared region although the accuracy specified in the catalogue was 20 MHz. This error was attributable to residual moisture inside the Fizeau interferometer of the wavelength meter. A simple method to avoid the error is proposed.