Synthetic MRI for Radiotherapy Planning for Brain and Prostate Cancers: Phantom Validation and Patient Evaluation.

Synthetic MRI for Radiotherapy Planning for Brain and Prostate Cancers: Phantom Validation and Patient Evaluation.
复制标题

DOI:
10.3389/fonc.2022.841761
复制
发表时间:
2022
影响因子:
4.7
通讯作者:
Gensanne, David
Gensanne, David
中科院分区:
医学3区
文献类型:
--
作者:
Gouel, Pierrick;Hapdey, Sebastien;Dumouchel, Arthur;Gardin, Isabelle;Torfeh, Eva;Hinault, Pauline;Vera, Pierre;Thureau, Sebastien;Gensanne, David

文献摘要

参考文献

被引文献

相似文献

We aimed to evaluate the accuracy of T1 and T2 mappings derived from a multispectral pulse sequence (magnetic resonance image compilation, MAGiC®) on 1.5-T MRI and with conventional sequences [gradient echo with variable flip angle (GRE-VFA) and multi-echo spin echo (ME-SE)] compared to the reference values for the purpose of radiotherapy treatment planning. The accuracy of T1 and T2 measurements was evaluated with 2 coils [head and neck unit (HNU) and BODY coils] on phantoms using descriptive statistics and Bland–Altman analysis. The reproducibility and repeatability of T1 and T2 measurements were performed on 15 sessions with the HNU coil. The T1 and T2 synthetic sequences obtained by both methods were evaluated according to quality assurance (QA) requirements for radiotherapy. T1 and T2 in vivo measurements of the brain or prostate tissues of two groups of five subjects were also compared. The phantom results showed good agreement (mean bias, 8.4%) between the two measurement methods for T1 values between 490 and 2,385 ms and T2 values between 25 and 400 ms. MAGiC® gave discordant results for T1 values below 220 ms (bias with the reference values, from 38% to 1,620%). T2 measurements were accurately estimated below 400 ms (mean bias, 8.5%) by both methods. The QA assessments are in agreement with the recommendations of imaging for contouring purposes for radiotherapy planning. On patient data of the brain and prostate, the measurements of T1 and T2 by the two quantitative MRI (qMRI) methods were comparable (max difference, <7%). This study shows that the accuracy, reproducibility, and repeatability of the multispectral pulse sequence (MAGiC®) were compatible with its use for radiotherapy treatment planning in a range of values corresponding to soft tissues. Even validated for brain imaging, MAGiC® could potentially be used for prostate qMRI.
DOI: 10.3938/jkps.74.892
发表时间: 2019-05-01
影响因子: 0.6
作者:
Kim, Seong-Ho;Park, Jin-Seo;Yang, Hyung-Jin
通讯作者: Yang, Hyung-Jin
DOI: 10.1186/s13550-019-0569-7
发表时间: 2019-12-10
期刊: EJNMMI RESEARCH
影响因子: 3.2
作者:
Koopman, Danielle;Jager, Pieter L.;van Dalen, Jorn A.
通讯作者: van Dalen, Jorn A.
DOI: 10.1259/bjr.73.868.10844863
发表时间: 2000-04-01
影响因子: 2.6
作者:
Firbank, MJ;Harrison, RM;Coulthard, A
通讯作者: Coulthard, A
DOI: 10.1007/s10334-016-0562-3
发表时间: 2016-10-01
影响因子: 2.3
作者:
Bojorquez, Jorge Zavala;Bricq, Stephanie;Lalande, Alain
通讯作者: Lalande, Alain
DOI: 10.1186/s12968-020-00613-3
发表时间: 2020-05-07
影响因子: 6.4
作者:
Captur, Gabriella;Bhandari, Abhiyan;Moon, James C.
通讯作者: Moon, James C.