Feasibility of measuring blood glucose concentration by near-infrared Raman spectroscopy.
Feasibility of measuring blood glucose concentration by near-infrared Raman spectroscopy.
复制标题
DOI:
10.1016/s1386-1425(96)01779-9
复制
发表时间:
1997-02
期刊:
影响因子:
--
通讯作者:
A. Berger;I. Itzkan;M. Feld
中科院分区:
文献类型:
--
作者:
A. Berger;I. Itzkan;M. Feld
We report the determinations of glucose concentration in human whole blood samples made using near-infrared Raman spectroscopy. Raman spectra of blood samples with above-physiological levels of glucose were acquired for 5 min through the wall of a cuvette via fiber optics. Partial least squares analysis was used to predict glucose concentrations in the samples. A root mean squared prediction error of 3.6 mM glucose was achieved with a correlation coefficient of 0.99 between reference and predicted values. This result is the first step in evaluating the potential of near-infrared Raman spectroscopy to perform blood glucose measurements with clinical accuracy. The technique is capable of measuring the concentration of other Raman-active blood constituents; as an example, bicarbonate was also measured. The method could eventually be useful for direct measurement of tissue analytes.