A case of variant hemoglobin (Hb Agenogi) with type 2 diabetes mellitus showed high HbA1c levels measured by immunoassay due to enhanced antigenicity
A case of variant hemoglobin (Hb Agenogi) with type 2 diabetes mellitus showed high HbA1c levels measured by immunoassay due to enhanced antigenicity
复制标题
一例患有 2 型糖尿病的变异血红蛋白 (Hb Agenogi) 病例由于抗原性增强,通过免疫测定测量显示 HbA1c 水平较高
DOI:
10.1007/s13340-018-0383-x
复制
发表时间:
2018
影响因子:
2.2
通讯作者:
M. Koga
中科院分区:
文献类型:
--
作者:
Y. Yamane;M. Tsumori;M. Ishibashi;M. Koga
HbA1c is a widely utilized biomarker for the management of diabetes mellitus. The presence of variant hemoglobins might interfere with HbA1c measurement using different methods. We herein describe Hb Agenogi (β90Glu → Lys) with type 2 diabetes mellitus whose HbA1c measured by immunoassay (IA) showed falsely high levels while HbA1c measured by high-performance liquid chromatography (HPLC) in the standard mode (SM) showed falsely low levels. To clarify the cause of the falsely high-IA-HbA1c levels, HbA1c was measured by various methods. Glycated albumin was slightly higher than the reference range. HbA1c measured by HPLC in the variant mode, affinity assay and enzymatic assay showed the range (mean ± 2SD: 6.0–6.8%) in this case. However, HbA1c measured by several HPLC-SMs showed falsely low levels (4.1–4.4%). IA-HbA1c using an antibody manufactured by Fujirebio Inc. showed falsely high levels (7.3–8.2%); whereas, IA-HbA1c using an antibody manufactured by Roche Ltd. showed the expected range (6.2–6.5%). In the case of Hb Agenogi, IA-HbA1c using an antibody manufactured by Fujirebio yielded falsely high levels. The mutation at codon 90 of the β-globin chain might enhance antigenicity of the N-terminal peptide region and, therefore, lead to falsely high-HbA1c levels in IA-HbA1c.
影响因子:
3.2
作者:
Committee of the Japan Diabetes Society on the Diagnostic Criteria of Diabetes Mellitus;Seino Y;Nanjo K;Tajima N;Kadowaki T;Kashiwagi A;Araki E;Ito C;Inagaki N;Iwamoto Y;Kasuga M;Hanafusa T;Haneda M;Ueki K
通讯作者:
Ueki K