Measurement and energy correction due to charge carrier lifetimes in large HPGe spectrometers
Measurement and energy correction due to charge carrier lifetimes in large HPGe spectrometers
复制标题
大型 HPGe 光谱仪中载流子寿命的测量和能量校正
DOI:
--
复制
发表时间:
2009
期刊:
影响因子:
--
通讯作者:
H. Miley
中科院分区:
文献类型:
--
作者:
J. Kephart;C. Aalseth;H. Miley
This work addresses the energy spectrum correction due to increased charge carrier collection times in larger HPGe spectrometers. The energy of the radiation interaction is expected to be proportional to the total collected charge. This is increasingly not true with larger HPGe spectrometers. Some charge is lost as the total charge travels from the interaction location to the collection electrode. This path dependent loss of charge results in decreased energy resolution. In HPGe spectrometers, this process is characterized by the charge carrier lifetime constant and is given as an exponential function of the charge carrier collection time divided by this constant. Thus large detectors can experience exponential decrease in energy resolution as charge carrier collection time increases. We studied the effect of charge carrier lifetime on energy resolution for a p-type point contact HPGe spectrometers using pulse shape analysis. We present a method using the rise time to correct for the charge carrier lifetime on a pulse by pulse basis for a given HPGe spectrometer.