Control of high common mode voltage during transthoracic defibrillation.
Control of high common mode voltage during transthoracic defibrillation.
复制标题
经胸除颤期间高共模电压的控制。
DOI:
10.1109/10.476128
复制
发表时间:
1995
期刊:
影响因子:
--
通讯作者:
Lerman,BB
中科院分区:
文献类型:
--
作者:
Deale,OC;Ng,KT;Kim,EJ;Lerman,BB
A high common mode voltage (V/sub cm/) relative to earth ground is produced on the myocardium during the delivery of a defibrillator pulse and can generate a differential error signal when potential gradients are recorded with bipolar electrodes and isolation amplifiers. The error signal is proportional to V/sub cm/, and therefore, a reduction in V/sub cm/ improves the accuracy of the potential gradient data. Experiments were conducted on 5 dogs to determine whether V/sub cm/ can be controlled using a bridge circuit. The bridge circuit consisted of a 5 k/spl Omega/ power rheostat in parallel with the transthoracic resistance of the dog. The variable contact of the rheostat was connected to earth ground, and by adjusting the rheostat, V/sub cm/ on the myocardium could be varied. In each dog, 20 A shocks were delivered through stainless steel transthoracic electrodes. Point contact electrodes sutured to the epicardium were used to measure V/sub cm/. It was determined that V/sub cm/ could be reduced to approximately zero at a given electrode on the heart. In addition, for the 5 dogs studied, the maximum measured V/sub cm/ on the heart was only 10% of the transthoracic voltage when the bridge circuit was balanced for an interior point in the heart.