OBJECTIVES
To determine the most appropriate sequences for the visualization of small parotid ducts in MR sialography.
METHODS
MR images of a phantom consisting of distilled water in polyethylene tubes were obtained with turbo-spin echo (TSE), single-shot turbo-spin echo (SSTSE), half-fourier acquisition, single-shot turbo-spin echo (HASTE) and turbo gradient-spin echo (TGSE) pulse sequences and compared visually and quantitatively. MR sialograms obtained from healthy volunteers and patients with Sjögren's syndrome (SS) were obtained using the same four sequences.
RESULTS
In the phantom, TSE images were best and the contrast-noise ratio (CNR) highest. In the volunteers, the main ducts were especially clearly visualized with TSE and in SSTSE; however, the majority of secondary and/or tertiary parotid ducts were not depicted by any of the sequences used. In SS patients, images of small main ducts and small pseudocysts were clearer using TSE. However, TSE could not depict the narrow main ducts or peripheral ducts or very small pseudocysts.
CONCLUSIONS
TSE is considered the most suitable MR sequence for assessing small parotid gland ducts. However, further improvement is needed since it does not always visualize them sufficiently.