Polarimetric Littrow Spectrograph - instrument calibration and first measurements
Polarimetric Littrow Spectrograph - instrument calibration and first measurements
复制标题
DOI:
10.1051/0004-6361:20052662
复制
发表时间:
2005-07
影响因子:
6.5
通讯作者:
C. Beck;W. Schmidt;T. Kentischer;D. Elmore
中科院分区:
文献类型:
--
作者:
C. Beck;W. Schmidt;T. Kentischer;D. Elmore
We present first measurements and the calibration procedure for the Polarimetric Littrow Spectrograph (POLIS) operated at the Vacuum Tower Telescope on Tenerife, together with a brief summary of the technical characteristics of the instrument. In its present configuration, we achieve a polarimetric accuracy of about $3 \times 10^{-3}$ in the visible channel (630 nm) of the instrument. The accuracy is limited by cross talk among the different polarization states. The detection limit for polarized light is about $2 \times 10^{-3}$ for a 7 s exposure. Polarimetric measurements in the blue channel ($\ion{Ca}{ii}$ H line, 396.7 nm) are strongly limited by the low photon flux. At this wavelength we present Stokes- V maps with a spatial resolution of about 0.5 arcs. The polarimetric quality of any spectropolarimeter is limited by the precision of the instrument calibration. We present a new method for self-calibration that reduces cross talk among the polarization components to 0.1%. This improvement results from a measurement of the retardance of the calibration waveplate with an accuracy of 0.1°. We demonstrate the capability of the simultaneous use of POLIS and the Tenerife Infrared Polarimeter which is integrated in the main spectrograph of the Vacuum Tower Telescope.