Measurement of glycated hemoglobin (HbA1c) with an automated POCT instrument in comparison with HPLC and automated immunochemistry method:: evaluation of the influence of hemoglobin variants
Measurement of glycated hemoglobin (HbA1c) with an automated POCT instrument in comparison with HPLC and automated immunochemistry method:: evaluation of the influence of hemoglobin variants
复制标题
DOI:
10.1515/cclm.2006.041
复制
发表时间:
2006-02-01
影响因子:
6.8
通讯作者:
Makris, K
中科院分区:
文献类型:
--
作者:
Haliassos, A;Drakopoulos, I;Makris, K
Our study evaluates a fast and easy way to perform point-of-care testing (POCT) measurements of glycated hemoglobin HbA(1c) in comparison with an immunoassay on an automated biochemistry analyzer and cation exchange chromatography, the two methods routinely used in clinical laboratories for the measurement of HbA(1c). A significant finding of our study is that although the POCT instrument insert claims that the method is not affected by the presence of HbS in the heterozygous state or in combination with beta-thalassemia, discrepant results were found in some cases with such hemoglobinopathies. In these cases, the two POCT and laboratory immunoassay methods showed clinically significant positive interferences with samples containing the HbS trait. We conclude that samples with the HbS trait should be interpreted with caution when tested using the POCT instrument.