Interlaboratory comparison of the Doumas bilirubin reference method
Interlaboratory comparison of the Doumas bilirubin reference method
复制标题
DOI:
10.1016/j.clinbiochem.2009.05.007
复制
发表时间:
2009-08-01
影响因子:
2.8
通讯作者:
Doumas, Basil T.
中科院分区:
文献类型:
--
作者:
Lo, Stanley F.;Kytzia, Hans-Joachim;Doumas, Basil T.
Objectives: To assess the performance of the Doumas bilirubin reference method.Design and methods: Ring trails using pooled patient specimens, a calibrator and human sera enriched with unconjugated bilirubin were analyzed in five laboratories using the Doumas bilirubin reference method.Results: The coefficient of variation for the linear measurement range between laboratories ranged from 1-3%.Conclusions: The Doumas bilirubin reference method is robust and reproducible. Bilirubin results using this method may be used in the development of more accurate and reliable calibrators. (C) 2009 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.