Measurement of thermal neutron fluence distribution with use of 23Na radioactivation around a medical compact cyclotron
Measurement of thermal neutron fluence distribution with use of 23Na radioactivation around a medical compact cyclotron
复制标题
利用医用紧凑型回旋加速器周围的 23Na 放射性激活测量热中子注量分布
DOI:
10.1007/s12194-009-0060-7
复制
发表时间:
2009
影响因子:
1.6
通讯作者:
T. Sakae
中科院分区:
文献类型:
--
作者:
T. Fujibuchi;I. Yamaguchi;Tetsuharu Kasahara;T. Iimori;Y. Masuda;K. Kimura;Hiroshi Watanabe;T. Isobe;T. Sakae
A medical compact cyclotron produces about 1015 neutrons per day along with 100 GBq of 18F. Therefore, it is important to establish radiation safety guidelines on residual radioactivity for routine operation, maintenance work, and decommissioning. Thus, we developed a simple method for measuring the thermal neutrons in a cyclotron room. In order to verify the feasibility of our proposed method, we measured the thermal neutron distribution around a cyclotron by using the activation of 23Na in salt. We installed 78 salt dosimeters in the cyclotron room with a 50 cm mesh. The photopeak of 24Na was measured, and the neutron flux distribution was estimated. Monitoring the neutron flux distribution in a cyclotron room appears to be useful for not only obtaining an accurate estimate of the distribution of induced radioactivity, but also optimizing the shield design for radiation safety in preparation for the decommissioning process.