Validation of a real-time bladder sensation meter during oral hydration in healthy adults: Repeatability and effects of fill rate and ultrasound probe pressure.
Validation of a real-time bladder sensation meter during oral hydration in healthy adults: Repeatability and effects of fill rate and ultrasound probe pressure.
复制标题
健康成人口服补水期间实时膀胱感觉计的验证:填充率和超声探头压力的重复性和影响。
DOI:
10.1111/luts.12275
复制
发表时间:
2019
期刊:
影响因子:
--
通讯作者:
Klausner,AdamP
中科院分区:
文献类型:
--
作者:
Sheen,Derek;Kolli,Hiren;Nagle,AnnaS;Cullingsworth,Zachary;Vinod,NaomiN;Naimi,HameedaA;DeWachter,StefanG;Mazzeo,SuzanneE;Stothers,Lynn;Speich,JohnE;Klausner,AdamP
ObjectivesA non‐invasive protocol was previously developed using three‐dimensional ultrasound and a sensation meter to characterize real‐time bladder sensation. This study the protocol by measuring the effects of fill rateand ultrasound probe pressure during oral hydration.MethodsHealthy volunteers with no urinary symptoms (based on International Consultation on Incontinence Questionnaire on Overactive Bladder surveys) were recruited into an oral hydration study. Throughout two complete fill–void cycles, participants drank 2 L Gatorade G2 (The Gatorade Company, Inc., Chicago, Illinois) and used a touch‐screen sensation meter to record real‐time bladder sensation (0%‐100%). The study was repeated three times, once per week (Visits A, B, and C). In Visits A and B, ultrasound was used to measure bladder volume every 5 minutes. Ultrasound was not used in Visit C except at 100% capacity. Volume data from Visit B were used to estimate volumes throughout the fills in Visit C. Sensation–capacity curves were generated for each fill for comparative analysis.ResultsTen participants completed three visits (60 total fills). Increased fill rate led to decreased sensation throughout filling, andultrasound probe pressure led to increased sensation. Participants reported higher sensation at low volumes during Fill 1 of Visit A before training with the sensation meter. Sensation curves with intermittent ultrasound showed repeatability for Fill 2 in Visits A and B. Fill rate and ultrasound probe pressure affect real‐time bladder sensation during oral hydration.ConclusionsThis study demonstrated repeatability of real‐time bladder sensation during a two‐fill oral hydration protocol with ultrasound.