Comparison of Dinamap 8100 with sphygmomanometer blood pressure measurement in a prepubertal diabetes cohort
Comparison of Dinamap 8100 with sphygmomanometer blood pressure measurement in a prepubertal diabetes cohort
复制标题
DOI:
10.1046/j.1440-1754.2001.00752.x
复制
发表时间:
2001-12-01
影响因子:
1.7
通讯作者:
Silink, M
中科院分区:
文献类型:
--
作者:
Jin, RZ;Donaghue, KC;Silink, M
Objective: To compare the conventional sphygmomanometer with the semiautomated Dinamap 8100 (Critikon, Tampa, FL, USA) for the measurement of blood pressure in prepubertal children with insulin-dependent diabetes mellitus.Methodology: Blood pressure was measured using both methods in 61 prepubertal children (aged 8-13 years) on 189 occasions over 4 years. The measurements were compared using the Bland-Altman plot. Tracking correlations of blood pressure centiles over time were analyzed by the general estimating equation.Results: Accuracy criteria of the Association for the Advancement of Medical Instrumentation were met and a British Hypertensive Society 'B' grading was reached. Differences in systolic and diastolic blood pressure were found between the two methods (P < 0.01). For systolic blood pressure, common correlations were 0.54 (Dinamap) and 0.51 (sphygmomanometer) and for diastolic blood pressure were 0.33 and 0.42, respectively.Conclusion: The Dinamap 8100 is an acceptable alternative in clinic practice and research for prepubertal children.