Determination of optimal angles for variable nutation proton magnetic spin-lattice, T1, and spin-spin, T2, relaxation times measurement
Determination of optimal angles for variable nutation proton magnetic spin-lattice, T1, and spin-spin, T2, relaxation times measurement
复制标题
DOI:
10.1002/mrm.10661
复制
发表时间:
2004-01-01
影响因子:
3.3
通讯作者:
Rutt, BK
中科院分区:
文献类型:
--
作者:
Deoni, SCL;Peters, TM;Rutt, BK
T-1 and T-2 can be rapidly determined with a combination of multiangle spoiled gradient recalled echo (SPGR) and steady-state free precession (SSFP) imaging. Previously, we demonstrated a simple method for determining the set of SPGR and SSFP angles that provided greater T-1 and T-2 precision than a set of uniformly spaced angles. In this article a more rigorous approach for determining angles is described. Weighted least-squares is also introduced for T-1 and T-2 estimation and a novel weighting function described. This new approach, suited for imaging applications where large T-1 and T-2 ranges are anticipated, provides high and uniform precision over a wide range of T-1 and T-2 values. (C) 2003 Wiley-Liss, Inc.