Combination of a low-tube-voltage technique with hybrid iterative reconstruction (iDose) algorithm at coronary computed tomographic angiography.
Combination of a low-tube-voltage technique with hybrid iterative reconstruction (iDose) algorithm at coronary computed tomographic angiography.
复制标题
DOI:
10.1097/rct.0b013e31821fee94
复制
发表时间:
2011-07
影响因子:
1.3
通讯作者:
Awai K
中科院分区:
文献类型:
--
作者:
Funama Y;Taguchi K;Utsunomiya D;Oda S;Yanaga Y;Yamashita Y;Awai K
We compare the performance of low-tube voltage with the hybrid iterative reconstruction (iDose) with standard- and low-tube voltage with the filtered backprojection (FBP) using phantoms at CT coronary angiography (CTCA). At CTCA, application of the combined low-tube voltage with iDose resulted in significant image quality improvements compared to the low-tube voltage with FBP. Image quality was the same or better despite a reduction in the radiation dose by 76% compared with standard-tube voltage with FBP.