Clinical validation of automated audiometry with continuous noise-monitoring in a clinically heterogeneous population outside a sound-treated environment
Clinical validation of automated audiometry with continuous noise-monitoring in a clinically heterogeneous population outside a sound-treated environment
复制标题
DOI:
10.1080/14992027.2016.1178858
复制
发表时间:
2016-01-01
影响因子:
2.7
通讯作者:
Atlas, Marcus D.
中科院分区:
文献类型:
--
作者:
Brennan-Jones, Christopher G.;Eikelboom, Robert H.;Atlas, Marcus D.
Objective: Examine the accuracy of automated audiometry in a clinically heterogeneous population of adults using the KUDUwave automated audiometer. Design: Prospective accuracy study. Manual audiometry was performed in a sound-treated room and automated audiometry was not conducted in a sound-treated environment. Study sample: 42 consecutively recruited participants from a tertiary otolaryngology department in Western Australia. Results: Absolute mean differences ranged between 5.12-9.68dB (air-conduction) and 8.26-15dB (bone-conduction). A total of 86.5% of manual and automated 4FAs were within 10dB (i.e.+/- 5dB); 94.8% were within 15dB. However, there were significant (p0.05; R-2=0.11). The presence of a hearing loss (better ear26dB) did not significantly affect accuracy (p=0.604; air-conduction), (p=0.218; bone-conduction). Conclusions: This study provides clinical validation of automated audiometry using the KUDUwave in a clinically heterogeneous population, without the use of a sound-treated environment. Whilst threshold variations were statistically significant, future research is needed to ascertain the clinical significance of such variation.