High-performance liquid chromatographic determination of theophylline and its derivatives with electrochemical detection.
High-performance liquid chromatographic determination of theophylline and its derivatives with electrochemical detection.
复制标题
电化学检测高效液相色谱法测定茶碱及其衍生物
DOI:
10.1016/0021-9673(75)85056-4
复制
发表时间:
1979
影响因子:
4.1
通讯作者:
W. J. Mayer
中科院分区:
文献类型:
--
作者:
M. Greenberg;W. J. Mayer
Theophylline and its derivatives are determined by high-performance liquid chromatographic separation on a C8reversed-phase column with electrochemical detection. The analyte is oxidized at 1.24 V (vs. the standard calomel electrode) on a wax-impregnated graphite electrode using classical d.c. and current sampled d.c. modes, and the results obtained compared with those of a 254 nm ultraviolet (UV) detector. Polarographic detection offers greater sensitivity for theophylline than UV detection; further, interferences which hamper UV detection may be selectively eliminated with polarographic detection.