Immunosensor using surface plasmon resonance for C-reactive protein detection

Immunosensor using surface plasmon resonance for C-reactive protein detection
复制标题

DOI:
10.5796/electrochemistry.74.153
复制
发表时间:
2006-02-01
期刊:
影响因子:
2.5
通讯作者:
Suzuki, H
Suzuki, H
中科院分区:
工程技术4区
文献类型:
--
作者:
Casa, E;Kurosawa, C;Suzuki, H

文献摘要

被引文献

相似文献

将便携式表面等离子体共振(SPR)仪与微流控SPR传感器芯片(MSSC)的抗人C-反应蛋白单克隆抗体(AbCRP)固定金表面相结合,研制了一种检测人C-反应蛋白(CRP)的免疫传感器。采用自组装技术,用4,4 '-二硫代二丁酸对MSSC金表面进行了化学修饰。随后,通过化学吸附将AbCRP固定在MSSC的化学活化金表面上,并使用免疫反应进行CRP测定。该开发的SPR方法对CRP浓度的检测限为1.0 μ g/ml。
An immunosensor for human C-reactive protein (CRP) detection was developed by combining a portable surface plasmon resonance (SPR) machine with an anti-human C-reactive protein monoclonal antibody (AbCRP)-immobilized gold surface of the microfluidic SPR sensor chip (MSSC). The gold surface of MSSC was chemically modified using self-assembly monolayer technique with 4,4'-dithiodibutyric acid. Subsequently, AbCRP was immobilized on the chemically-activated gold surface of the MSSC by chemical adsorption and CRP determinations were performed using immunoreaction. The detection limit for CRP concentration by this developed SPR method was 1.0 mu g/ml.